Tubestand C

TS9611
The Tubestand C has been designed to support and move a radiogen assembly (RX tube and diaphragm).

With the stand, it is possible to position the radiogen assembly as so to direct the radiant beam towards receptor devices such as a fixed or elevating and tilting table top bucky table, a stretcher, a wall bucky stand, a cassette changer, etc.

Movements are manual and the mechanically balanced mobile parts are kept in parking position by electromagnetic brakes. For the movements, in-line electromagnetic brakes are used while, for the tube rotation movement on the horizontal arm of support, brakes at permanent magnet are used that assures the braking action even in case of a sudden lack of current.

The equipment is made with aluminium profile special alloy. The equipment can be easily checked in each part so maintenance and inspections to keep its performance, safe use and reliability unchanged in the long run are made easier.

Please contact us by phone 01902 637333 or email sales@wolversonx-ray.co.uk for an accurate quotation.

Tubestand C floor/ceiling tube column stand with magnetic locks.Spring counterpoised vertical movement. Transverse cross arm movement +/- 11 cm. Complete with light beam collimator.

Longitudinal rail

  • Length 400 cm
  • Longitudinal travel 280 cm

Column rotation

  • Angle of rotation +/- 90° with mechanical detent in the orthogonal positions
  • Unlocking rotational movement foot pedal

Vertical

  • Focus to floor minimum distance 48 cm
  • Vertical travel 150 cm
  • Focus to film indicator Adhesive scaled overlay (metric system)
  • Balancing system counterweights

Transversal

  • Focus to column vertical axis distance 80 cm
  • Transversal movement +/- 11 cm

Tube horizontal rotation

  • Angle of rotation +/- 180° with mechanical detent in the orthogonal positions
  • Angle indicator goniometer
  • Locking of rotation In every position with magnetic brake

Click here to download the Tubestand C brochure